Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00254-1 |
Product Name/Description |
VITROS Chemistry Products PHBR Slides (An in vitro diagnostic medical device (IVD) for the quantitative measure of phenobarbital)
GEN 38. Coating 0060 GEN 39. Coating 0062 GEN 40. Coating 0063 GEN 41. Coating 0064
Expiry date range: 1 April 2013 through 1 February 2014 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
19/03/2013 |
Responsible Entity |
|
Reason/Issue |
Ortho Clinical Diagnostics has received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. An investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids. |
Recall Action |
Recall |
Recall Action Instructions |
End users are requested to discard all lots of VITROS PHBR Slides. It is recommended that any concerns with previously reported patient results be discussed with the laboratory director and requesting physician. |
Contact Information |
1800 032 359 - Ortho-Clinical Diagnostics Customer Technical Services |