Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00237-1
Product Name/Description Avance CS2 Anesthesia Carestation

All devices are affected

ARTG Number: 93955
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 15/03/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue The Avance CS2 may experience an intermittent display unit failure while in use, which could result in unexpected system performance and may include the loss of mechanical ventilation and balance gas delivery (Medical AIR or N20)
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising end users of work around instructions to mitigate safety issues. A correction is planned to be implemented for all affected systems.
Contact Information 1300 722 229 - GE National Service and Support Centre