Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00188-1
Product Name/Description All Plum A+ Family of Infusers: Plum A+Single Channel Infusion Pump, Plum A+ and A+3 Infusion Pumps with MedNet Software

Plum A+Single Channel Infusion Pump
Catalogue Number: 611239101
Number on Pump: 12391 & 11971

Plum A+ Infusion Pump with MedNet Software
Catalogue Number: 612079201
Number on Pump: 20792

Plum A+3 Infusion Pump with MedNet Software
Catalogue Number: 612067801
Number on Pump: 20678
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 27/02/2013
Responsible Entity Hospira Australia Pty Ltd
Reason/Issue The distal pressure sensor calibration may have drifted and require re-calibration. The following error codes may be displayed during set up or infusion on the affected devices: E180/N180, E181/N181, E186/N186, E187/N187, E346. If distal pressure sensor drift occurs, the pump may not sense the build up of pressure and will not alarm when occlusion thresholds are exceeded.
Recall Action Recall for Product Correction
Recall Action Instructions Hospira is providing end users with instructions to determine if they are affected by the issue. Healthcare professionals are advised to weigh the risk/benefit to patients associated with the use of the device when administering critical therapies. Hospira is also in the process of evaluating the design of the distal pressure sensor to reduce the opportunity for pressure sensor drift. Additionally, a yearly test to determine if the distal pressure sensor’s calibration has drifted will be released in 2013.
Contact Information 1300 046 774 (Option 1) - Hospira Customer Service