Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00188-1 |
Product Name/Description |
All Plum A+ Family of Infusers: Plum A+Single Channel Infusion Pump, Plum A+ and A+3 Infusion Pumps with MedNet Software
Plum A+Single Channel Infusion Pump Catalogue Number: 611239101 Number on Pump: 12391 & 11971
Plum A+ Infusion Pump with MedNet Software Catalogue Number: 612079201 Number on Pump: 20792
Plum A+3 Infusion Pump with MedNet Software Catalogue Number: 612067801 Number on Pump: 20678
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
27/02/2013 |
Responsible Entity |
|
Reason/Issue |
The distal pressure sensor calibration may have drifted and require re-calibration. The following error codes may be displayed during set up or infusion on the affected devices: E180/N180, E181/N181, E186/N186, E187/N187, E346. If distal pressure sensor drift occurs, the pump may not sense the build up of pressure and will not alarm when occlusion thresholds are exceeded. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Hospira is providing end users with instructions to determine if they are affected by the issue. Healthcare professionals are advised to weigh the risk/benefit to patients associated with the use of the device when administering critical therapies. Hospira is also in the process of evaluating the design of the distal pressure sensor to reduce the opportunity for pressure sensor drift. Additionally, a yearly test to determine if the distal pressure sensor’s calibration has drifted will be released in 2013. |
Contact Information |
1300 046 774 (Option 1) - Hospira Customer Service |