Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00186-1
Product Name/Description BiCision (electrosurgical thermofusion and dissection system)

Lot / serial numbers:
W3174973
WO182954
W3177733
W3177734
W3179128
WN189444
W3176581
WN189947

ARTG Number: 94173
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/02/2013
Responsible Entity Rymed Pty Ltd
Reason/Issue Internal investigations found very few Bicision instruments with the quality issue of low closing force of the jaw, which under unfavourable conditions could lead to insufficient thermofusion of large vessels or tissue bundles.

This recall action was carried out prior to review by the TGA.
Recall Action Recall
Recall Action Instructions Rymed is advising users to remove all affected devices from use. Return all affected devices to Rymed for credit or replacement when available.
Contact Information 02 9676 3488 - Rymed