Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00182-1
Product Name/Description RaySearch RayStation 3.0 (Radiation treatment planning and analysis software)

Version 3.0.0.251

ARTG Number: 195288
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/02/2013
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions.
Recall Action Recall for Product Correction
Recall Action Instructions Emergo is providing users with a work around to mitigate the issue until a corrective action can be implemented. A software update will be provided to correct the issue as soon as possible.
Contact Information 02 9006 1662 - Emergo Australia