Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00182-1 |
Product Name/Description |
RaySearch RayStation 3.0 (Radiation treatment planning and analysis software)
Version 3.0.0.251
ARTG Number: 195288 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
26/02/2013 |
Responsible Entity |
|
Reason/Issue |
Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Emergo is providing users with a work around to mitigate the issue until a corrective action can be implemented. A software update will be provided to correct the issue as soon as possible. |