Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00171-1
Product Name/Description Drug 4 Cal (KC460) & Drug 4 Cal (KC460A) for use with the Dimension Vista System. In vitro diagnostic medical devices (IVD).

Vista Drug 4 Calibrator
Catalogue number: KC460
Siemens material number: 10472284
Lot number: 2KD052

Vista Drug 4 Calibrator Lot (SMN )
Catalogue number: KC460A
Siemens material number: 10476171
Lot number: 2KD053

ARTG Number: 182220
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/02/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista® LOCI Digoxin (DIGXN) Flex® reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL.
Recall Action Recall
Recall Action Instructions Customers are requested to immediately discard any remaining stock. Siemens is asking customers to discuss the issues with the laboratory director so that they can determine if additional follow-up is required for patient tests conducted with the affected lots.
Contact Information 1800 310 300 - Siemens Technical Support