Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00156-1
Product Name/Description Philips HeartStart XL+ Defibrillator / Monitor (manual defibrillator)

Model Number: 861290

Serial Number ranges:
USO1100100–USO1100372
US61201722–US61201924
USN1100376–USN1100960
US71201925–US71202048
USD1100961–USD1101095
US81202049–US81202168
US11201096–US11201186
US91202169–US91202514
US21201187–US21201239
USO1202515–USO1202990
US31201240–US31201537
USN1202991–USN1203537
US41201538–US41201585
USD1203538–USD1203968
US51201586–US51201721
US11303969–US11303972
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 19/02/2013
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue The Philips HeartStart XL+ Defibrillator/Monitor may fail to power on in clinical mode and instead power on in service mode, displaying an "Equipment Disabled: Therapy" prompt, requiring the user to enter a password and perform an Operational Check to clear the condition. The device will not return to clinical mode until an Operational Check has been performed.
Recall Action Recall for Product Correction
Recall Action Instructions A Philips representative with contact customers with affected devices to arrange for a software update. Until the correction is implemented the device can continue to be used by following work around procedures.
Contact Information 1800 251 400 - Philips Customer Care Centre