Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00156-1 |
Product Name/Description |
Philips HeartStart XL+ Defibrillator / Monitor (manual defibrillator)
Model Number: 861290
Serial Number ranges: USO1100100–USO1100372 US61201722–US61201924 USN1100376–USN1100960 US71201925–US71202048 USD1100961–USD1101095 US81202049–US81202168 US11201096–US11201186 US91202169–US91202514 US21201187–US21201239 USO1202515–USO1202990 US31201240–US31201537 USN1202991–USN1203537 US41201538–US41201585 USD1203538–USD1203968 US51201586–US51201721 US11303969–US11303972 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
19/02/2013 |
Responsible Entity |
|
Reason/Issue |
The Philips HeartStart XL+ Defibrillator/Monitor may fail to power on in clinical mode and instead power on in service mode, displaying an "Equipment Disabled: Therapy" prompt, requiring the user to enter a password and perform an Operational Check to clear the condition. The device will not return to clinical mode until an Operational Check has been performed. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
A Philips representative with contact customers with affected devices to arrange for a software update. Until the correction is implemented the device can continue to be used by following work around procedures. |
Contact Information |
1800 251 400 - Philips Customer Care Centre |