Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00149-1
Product Name/Description KWIK-QC Acid Fast Stain Slides (Microbiology quality control slides). An in vitro diagnostic medical device (IVD)

Product Code: SL42-10
Lot Number: 4242
Expiry Date: 20/06/2014
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 15/02/2013
Responsible Entity Abacus ALS Pty Ltd
Reason/Issue Control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain.
Recall Action Recall
Recall Action Instructions Abacus ALS is requesting customers supplied with the affected lot to destroy all remaining stock. Abacus will arrange for replacement with unaffected stock.