Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00147-1
Product Name/Description Vitrea Enterprise Suite : Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX (Tool for analysis of diagnostic images)

All versions prior to 6.3.3, except 6.1.6, 6.2.3 and 6.3 are affected

ARTG Number: 154823
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 15/02/2013
Responsible Entity Toshiba Australia Pty Ltd
Reason/Issue During internal testing, the manufacturer Vital images has found that an error can occur when images are rotated in non-90 degree increments of the transverse/axial plane (rotation around the z-axis), while not being simultaneously rotated around the x-axis and y-axis. This orientation adjustment has been found to occur on rare occasions, such as to straighten tilts in head or neck scans, usually with minimal effect on measurements, but sometimes resulting in differences in length measurements of up to twenty nine percent and in volume measurements of up to fifty percent.
Recall Action Recall for Product Correction
Recall Action Instructions Vital Images recommends that all measurements associated with the issue are verified against results from other technologies. A software patch is under development to correct the issue for software versions 5.2, 6.0,6.1,6.2 and 6.3.
Contact Information 1300 166 155 - Toshiba Response Centre