Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00129-1
Product Name/Description SureSigns Patient Monitors, Model VS2+

Product number: 863278, 863279

Multiple affected serial numbers

ARTG Number: 93971
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 11/02/2013
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips Healthcare has identified that it is possible for the SureSigns VS2+ Patient Monitor to lock up and stop monitoring during patient use.
Recall Action Recall for Product Correction
Recall Action Instructions Philips is advising end users to not use the device in non-invasive blood pressure (NBP) Interval mode. If the device freezes disconnect the power and remove the battery. Reconnect the battery and power-up after 60 seconds.

Philips Healthcare is implementing a design change to correct the issue.
Contact Information 1800 251 400 - Philips Customer Care Centre