Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00111-1
Product Name/Description Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos RHD, Luminos RF Classic, Luminos Fusion, Luminos Select (Fluoroscopic diagnostic x-ray systems)

ARTG Numbers: 102181 & 144246
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/02/2013
Responsible Entity Siemens Ltd
Reason/Issue There is potential for malfunction in cases where the table vertical tilt movement is obstructed by a solid object resulting in a collision. This can result in damage to the system mechanics that may not necessarily prevent the table from functioning immediately, but can lead to complete failure of the table mounting mechanics. In this case, the table may drop resulting in potentially hazardous situations for patients and/or staff.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising customers to avoid any collision of the table with objects in the table's operating area. Customers should contact their Siemens service representative for a precautionary inspection should such a collision occur or have occurred in the past.
Contact Information 1800 310 300 - Siemens Regulatory Affairs