Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00107-1
Product Name/Description UniCel DxI 800 Access Immunoassay Systems, Unicel DxI 600 Access Immunoassay Systems, UniCel DxC 660i, 680i, 860i, 880i SYNCHRON Access Clinical System. In vitro diagnostic medical devices (IVDs)

Part numbers: 973100, A30260, A71456, A71457, A71460, A71461, A25285, A25288, A92060
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/02/2013
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Improper tubing installing by the customer during routine maintenance may prevent the DxI and DxC systems from washing reaction vessels properly and may lead to erroneous test results. The reason for the customer mis-routing of instrument tubing was determined to be two fold:
- The mechanical controls on the instrument were insufficient to prevent incorrect routing of the aspirate and dispense tubing.
- The instructional drawings in the service documentation were difficult for the customer to understand.

This recall action was not notified to the TGA before it was initiated by Beckman Coulter Australia
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter Australia is advising that all affected systems were corrected with a new design that provided mechanical controls to reduce the potential for tubing installation errors. Also, updated service documentation with clearer instructions were distributed to affected customers.
Contact Information 1800 060 881 - Beckman Coulter Customer Support