Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00091-2 |
Product Name/Description |
Level 1 Fast Flow Fluid Warmers with and without Air Detection
Item Code: H-1200 Affected Serial Numbers: S103A00028, S103A00029, S103A00094 and S103A00095 ARTG Number: 108420 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
31/01/2013 |
Responsible Entity |
|
Reason/Issue |
A limited number of fluid warmers were manufactured with an incorrect printed circuit board component which could prevent the fluid warmer from functioning as intended and also potentially exhibiting a false over temperature alarm. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Smiths Medical Australasia is requesting users to identify any affected units and place in quarantine. A Biomedical Technician will inspect devices to determine if they are affected and provide a correction. |
Contact Information |
1800 654 949 - Smiths Medical Customer Service Department |