Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00091-2
Product Name/Description Level 1 Fast Flow Fluid Warmers with and without Air Detection

Item Code: H-1200
Affected Serial Numbers: S103A00028, S103A00029, S103A00094 and S103A00095
ARTG Number: 108420
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 31/01/2013
Responsible Entity Smiths Medical Australasia Pty Ltd
Reason/Issue A limited number of fluid warmers were manufactured with an incorrect printed circuit board component which could prevent the fluid warmer from functioning as intended and also potentially exhibiting a false over temperature alarm.
Recall Action Recall for Product Correction
Recall Action Instructions Smiths Medical Australasia is requesting users to identify any affected units and place in quarantine. A Biomedical Technician will inspect devices to determine if they are affected and provide a correction.
Contact Information 1800 654 949 - Smiths Medical Customer Service Department