Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00087-2
Product Name/Description HeartStart Automated External Defibrillator (AED)

Models affected : FRx and HS1
All serial numbers affected
ARTG Number: 95661
Recall Action Level Retail
Recall Action Classification Class I
Recall Action Commencement Date 31/01/2013
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips is providing additional instructions for use to advise the user of the correct actions to take in the event an error notification:

-If the FRx or HS1 AED emits a pattern of 'single chirps' please press the blue i-button on the front of the device. The AED will tell exactly what actions
to take to ensure that the device is ready for use.

-If at any time during the life of the device, the FRx or HS1 AED emitted or begins to emit a pattern of 'triple chirps', it is important that the
device is removed from use and a Philips representative is contacted. It is a signal that the device requires investigation by Philips to ensure that it is
ready for use.
Recall Action Recall for Product Correction
Recall Action Instructions Philips is providing additional instructions advising users to contact Philips technical support if a device error occurs and the device begins to emit a triple chirp, even if the error can be cleared.
Contact Information 1800 251 400 - Philips Customer Care Centre