TGA Recall Reference RC-2013-RN-00069-1
Product Name/Description Mizuho OSI 5855 Orthopedic Table Top

Manufactured January 1996 through December 2007

ARTG 183118
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/01/2013
Responsible Entity Life Healthcare Pty Ltd
Reason/Issue This action has been initiated to eliminate the risk of de-lamination of the epoxy bond between the 5855-01 Orthopedic Table Spar's aluminium insert and the carbon fibre spar itself. If the epoxy bond fails the aluminium plug may migrate out of the carbon fibre tube.
Recall Action Recall for Product Correction
Recall Action Instructions Customers are being advised to inspect the table top for any evidence of a shift of the aluminium plug from the carbon spar and immediately contact LifeHealthcare if apparent. Hospitals will be contacted to schedule an appointment to replace the table spar.