Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00067-1 |
Product Name/Description |
Medtronic Ardian Symplicity Catheter System
Model RDN 006
All lot numbers are affected
ARTG Number: 198985 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/01/2013 |
Responsible Entity |
|
Reason/Issue |
Medtronic has observed a connection issue with the Medtronic Ardian Symplicity Catheter System that can cause unstable temperature readings or lack of signal to the generator. In events where a connection failure leads to unstable temperature reading, the generator's safety algorithm is triggered and energy delivery is terminated, or if there is a lack of signal to the generator, it will not initiate treatment. |
Recall Action |
Recall |
Recall Action Instructions |
If the connection issue occurs the user is advised to assure that the catheter connection plug is fully seated into the generator by unplugging and re-inserting or gently wiggling the connector plug. If the connection issue is not resolved replace the catheter and contact Medtronic to organise return of the device. |
Contact Information |
02 9888 1283 - Medtronic Regulatory Affairs |