Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00067-1
Product Name/Description Medtronic Ardian Symplicity Catheter System

Model RDN 006

All lot numbers are affected

ARTG Number: 198985
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/01/2013
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue Medtronic has observed a connection issue with the Medtronic Ardian Symplicity Catheter System that can cause unstable temperature readings or lack of signal to the generator. In events where a connection failure leads to unstable temperature reading, the generator's safety algorithm is triggered and energy delivery is terminated, or if there is a lack of signal to the generator, it will not initiate treatment.
Recall Action Recall
Recall Action Instructions If the connection issue occurs the user is advised to assure that the catheter connection plug is fully seated into the generator by unplugging and re-inserting or gently wiggling the connector plug. If the connection issue is not resolved replace the catheter and contact Medtronic to organise return of the device.
Contact Information 02 9888 1283 - Medtronic Regulatory Affairs