Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00046-1 |
Product Name/Description |
CEM Nosecone for use with CUSA Excel/Excel+ Ultrasonic Aspirator
Catalogue Numbers: C6623, C6636
All lot numbers affected
ARTG Number: 197945 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
17/01/2013 |
Responsible Entity |
|
Reason/Issue |
Integra Neurosciences has been notified by the manufacturer that since 2009 they have received a small number of complaints for CUSA tip breakage. The manufacturer's investigation of these complaints identified that when the tips are used under specific circumstances with the CEM Nosecone, there is the potential for erosion of the CUSA tip. The erosion can potentially lead to tip breakage. Specifically, the potential for CUSA tip erosion occurs only during CEM operation and if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Integra Neurosciences Pty Ltd is revising the package insert to include an additional warning advising users to ensure that only the CEM Nosecone contacts the intended tissue at the annulus of the tip and not the side of the tip to avoid tip breakage. |
Contact Information |
1300 550 599 - Integra Customer Service |