Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00031-1
Product Name/Description BM Lactate Test Strips for use with the Accutrend Plus, Accutrend GC/GCT/Lactate. An in vitro diagnostic medical device (IVD)

All lots (2133260x, 2153300x, 2173380x) within shelf life are affected

Material Numbers: 3012654016 & 3012654370

ARTG Number: 201876
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/01/2013
Responsible Entity Roche Diagnostics Australia Pty Limited
Reason/Issue The BM-Lactate test strips show a positive bias of 1.0 to 1.5 mmol/L with patient samples.
Recall Action Recall
Recall Action Instructions Roche Diagnostics Australia Pty Ltd is requesting that customers cease use and discard all lots of Accutrend BM-Lactate.
Contact Information 1800 645 619 - Roche National Support Centre