Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00025-1
Product Name/Description HAMILTON-T1 Transport Ventilator, Software version 1.1.2

Model Number: T1
Multiple serial numbers affected
ARTG Number: 187874
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 10/01/2013
Responsible Entity AHC Solutions Pty Ltd
Reason/Issue Higher than expected internal oxygen consumption of the Hamilton T1 transport ventilator may cause miscalculation of the required oxygen supply for paediatric patient transport situations with long timeframe applications where the oxygen supply may be limited or finite.
Recall Action Recall for Product Correction
Recall Action Instructions AHC Solutions Pty Ltd (the sponsor) is providing amended instructions describing the process required to calculate the correct oxygen requirement prior to transport. Hamilton Medical (the manufacturer) will provide a software update to correct the issue.