Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00023-1
Product Name/Description Giraffe Incubator and Giraffe OmniBed with Servo Oxygen Control System

Manufactured between August 1 2012 and December 20 2012

Multiple serial numbers affected

ARTG Number: 139289
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/01/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue When the giraffe OmniBed and Giraffe Incubator is powered "on" or there is a recovery from power failure, the oxygen set point or radiant heater set point (values that are entered by the clinician) may be overwritten due to a potential overwriting of stored information associated with the display printed circuit board.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising hospitals may continue to use the affect product, however, all set points must be confirmed when powering 'on' the Giraffe microenvironment or when recovering from a power failure event. A software update will be provided by GE Healthcare to correct the issue.
Contact Information 1300 722 229 - GE National Service and Support Centre