Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00012-1
Product Name/Description MX 16-Slice CT scanner

Serial numbers: EP16E090006, EP16E110104, EP16E100082, EP16E120011, P16EM09002, EP16E100003

ARTG Number: 158775
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/01/2013
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips discovered that four screws that secure the patient table top to the carrier can become loose, causing the table top to detach.
Recall Action Recall for Product Correction
Recall Action Instructions Phillips Healthcare is advising hospitals and radiology centres to inspect the patient table top screws. If the screws are not properly seated, discontinue use of the device and contact a Philips Field Service Engineer for repair. Philips Healthcare will contact their customers to schedule a time to replace the screws for all affected systems.
Contact Information 1800 251 400 - Philips Customer Care Centre