Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01292-1 |
Product Name/Description |
ADVIA Centaur and ADVIA Centaur XP HBsAg II Assay. An in vitro diagnostic medical device (IVD)
ADVIA Centaur and ADVIA Centaur XP HBsAg II (HBsII) Assay
Reference Number: 10492138
Kit Lot numbers: 93076013, 94615013, 95344013, 95801013, 96628013, 98117014, 98508014, 12536014
An in vitro diagnostic medical device (IVD) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
24/12/2012 |
Responsible Entity |
|
Reason/Issue |
There is an increased number of falsely elevated patient samples for HBsAgII resulting in the number of initial positives reactive rates exceeding the initial reactive specificity as listed in the Instructions for Use (IFU). The increased false reactive rate may cause a delay in reporting results because additional testing is required. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is providing work around instructions for users to implement. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |