Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01292-1
Product Name/Description ADVIA Centaur and ADVIA Centaur XP HBsAg II Assay. An in vitro diagnostic medical device (IVD)

ADVIA Centaur and ADVIA Centaur XP HBsAg II (HBsII) Assay

Reference Number: 10492138

Kit Lot numbers: 93076013, 94615013, 95344013, 95801013, 96628013, 98117014, 98508014, 12536014

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/12/2012
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue There is an increased number of falsely elevated patient samples for HBsAgII resulting in the number of initial positives reactive rates exceeding the initial reactive specificity as listed in the Instructions for Use (IFU). The increased false reactive rate may cause a delay in reporting results because additional testing is required.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing work around instructions for users to implement.
Contact Information 1800 310 300 - Siemens Technical Support Centre