Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01289-1
Product Name/Description Dimension Clinical Chemistry System, Total Iron Binding Capacity. An in vitro diagnostic medical device (IVD)

Test Code: IBCT
Catalogue Number: DF84
Siemens Material Number: 10444944
ARTG number: 181687

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 20/12/2012
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Through monitoring of method performance Siemens Healthcare Diagnostics has determined the analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is provideding work around instructions for users to implement.
Contact Information 1800 310 300 - Siemens Technical Support Centre