Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01285-1
Product Name/Description Animas 2020 Insulin Infusion Pumps (ambulatory infusion pump)

ARTG Number: 96120
Recall Action Level Consumer
Recall Action Classification Class I
Recall Action Commencement Date 19/12/2012
Responsible Entity Australasian Medical & Scientific Ltd
Reason/Issue There are a small number of units where the U4 component on the printed circuit board is misaligned potentially causing the pump to dispense all of the insulin in the cartridge.
Recall Action Recall for Product Correction
Recall Action Instructions All users of the Animas 2020 Insulin Pump are receiving communication from the sponsor with work around instructions and instructions on how to return the affected pumps for a replacement pump
Contact Information 1300 851 056 - Animas / AMSL Patients Support Line