Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01255-1
Product Name/Description AU5800 Clinical Chemistry Analyzer, software versions 4.05 and prior. An in vitro diagnostic medical device (IVD)

All serial numbers affected

All software versions 4.05 and prior are affected.

An in vitro diagnostic medical device
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/12/2012
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is providing updated work instructions to mitigate the issue. A software update will be implemented as a permanent solution.
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre