Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01235-1
Product Name/Description ARIA Radiation Oncology System (information system), versions: 11.0 below 11.0.55, Build 11.02.28 and 11.0.34

Product code: HIT

Affected ARIA versions: 11.0 below 11.0.55, Build 11.02.28 and 11.0.34

Affected serial numbers: HIT0339M, HIT0339R1 & HIT4909

ARTG number: 132638
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 7/12/2012
Responsible Entity Varian Medical Systems Australasia Pty Ltd
Reason/Issue An anomaly has been identified during internal investigation with the ARIA Oncology Information System version 11 DICOM Import/Export functionality where a 360 degree arc field may convert to static field during DICOM export or import.
Recall Action Recall for Product Correction
Recall Action Instructions Varian is providing work around instructions to users and is developing a permanent solution.
Contact Information 1800 657 036 - Varian Oncology Helpdesk