Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01227-1 |
Product Name/Description |
AMH Gen II ELISA (clinical chemistry substrate, Anti-Mullerian Hormone assay). An in vitro diagnostic medical device (IVD).
Part number: A79765
All lot numbers are affected
ARTG number: 185654 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
5/12/2012 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has confirmed through customer feedback and internal investigation that some samples diluted with the AMH Gen II Sample Diluent prior to running the assay may generate results that are falsely elevated. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter is providing updated work around instructions to mitigate the issue and asking users to notify clinicians/research scientists regarding the potentially affected results at the discretion of the Laboratory Director. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |