Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01222-1
Product Name/Description DualHemo Mcable for use with the Infinity Acute Care System (IACS) (Patient monitoring system)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/12/2012
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue There have been reported cases in which the displayed value for the invasive blood pressure measurement of the IACS was lower than the actual patient
values were. Investigations have shown that liquid ingress in the cable connection of the DualHemo Mcable may cause inaccurate measurements.
Further investigation revealed that this can happen if the DualHemo Mcable is used in a non-vertical position or vertically but top down instead of top up, or if it
is simply laid down on the floor or on the patient bed and is not adequately positioned.
Recall Action Recall for Product Correction
Recall Action Instructions Drager Medical GmbH is providing an amendment to the instructions for use describing the correct orientation of the DualHemo Mcable when in use.