Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01202-1 |
Product Name/Description |
DBB-06, DBB-07 Haemodialysis Device (haemofiltration dialysis unit)
Reference Numbers: GE-F110-00, GE-F111-00, GE-F112-00, GE-F114-00 & GE-F115-00
All devices with serial numbers manufactured prior to 1 October 2012
ARTG Number: 189947 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
6/12/2012 |
Responsible Entity |
|
Reason/Issue |
NIKKISO has received complaints regarding too high a patient leakage current with the DBB-06 devices, used in combination with a central venous catheter (CVC). These complaints were related to technical measurement without any patient involvement. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Baxter Healthcare is recommending work around instructions to mitigate the risk. A modification to the hardware is planned to correct the issue. |
Contact Information |
02 8845 1601 - Baxter Quality Assurance |