Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01199-1
Product Name/Description Clinical Chemistry Phenobarbital Reagent. An in vitro diagnostic medical device (IVD)

List Number: 1E08

Lot Numbers: 52803UN12 (expired), 62299UN12 & 85773UN12

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/11/2012
Responsible Entity Abbott Australasia Pty Ltd Diagnostic Division
Reason/Issue Abbott Diagnostics is recalling this product due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital is 'less than or equal to' 7% Total CV%. Internal testing has confirmed CVs up to approximately 12% for in-date reagent lots over five months old.
Recall Action Recall
Recall Action Instructions Abbott is asking users to discontinue use and dispose of the affected units.
Contact Information 1800 816 696 - Abbott Customer Support