Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01179-1 |
Product Name/Description |
Phenobarbitol Reagent (used with ADVIA Chemistry System). An in vitro diagnostic medical device (IVD)
Phenobarbitol Reagent (PHNB_2)
Catalogue Number: 10377505
Lot Numbers: 244574, 256536
ARTG Number: 176643
An in vitro diagnostic medical device (IVD) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
28/11/2012 |
Responsible Entity |
|
Reason/Issue |
Internal studies have indicated there is an error which may cause control values and/or patient values to be increased or decreased by as much as 30%. |
Recall Action |
Recall |
Recall Action Instructions |
Siemens Healthcare is asking customers to discontinue use and dispose the affected stock. A review of previously generated results should be undertaken at the discretion of the Laboratory Director. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |