Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01161-1
Product Name/Description A3/A5 Anesthesia Delivery System

ARTG:158607
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 21/12/2012
Responsible Entity Ulco Engineering Pty Ltd
Reason/Issue There is the possibility of a system leak resulting from improper seating of the CO2 absorbent canister gasket. Should a system leak occur during use, fresh gas flow may be increased to compensate for any pressure loss resulting from the leak while the absorbent canister is unlocked, and locked again to reseat the gasket. If the CO2 absorbent gasket is improperly seated, however a potential leak will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case, at which time the absorbent canister can be unlocked and locked again to reseat the gasket.

This recall action was not notified to the TGA before it was initiated by Ulco Medical.
Recall Action Recall for Product Correction
Recall Action Instructions Ulco Medical is providing work around instructions and will be replacing the canister gasket to permanently correct the issue.
Contact Information 02 9519 5881 - Ulco Medical