Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01151-1
Product Name/Description YUNO Extension Device (used to support and stabilise the patient's body during a surgical procedure)

Part Number: 1433.62A1; Serial number: 001-107

Part Number: 1433.62A0; All serial numbers

ARTG Number: 163023
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/11/2012
Responsible Entity Maquet Australia Pty Ltd
Reason/Issue Maquet has identified a potential issue with the release lever of the screw tension mechanism, whereby it may be difficult to set or maintain the desired traction. This issue can occur during use with any patient population, however it is most likely to occur when setting higher tensions.
Recall Action Recall for Product Correction
Recall Action Instructions A Maquet authorised Service Engineer is to arrange an appointment to assess the affected devices and to replace the release lever. Customers are advised to be aware that they may experience difficulty with the tension mechanism until the repair/replacement activity is performed.