Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01145-1 |
Product Name/Description |
Steris Reliance 130L Cart Washer (decontamination washer utensil)
Product Code: ST-FW-0220321101
Manufactured between March 1999 and October 2009
ARTG Number: 135530 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/11/2012 |
Responsible Entity |
|
Reason/Issue |
Steris has received reports of the washer door not functioning as expected. To ensure end users are adequately protected from potential exposure to hot water/steam a 'door close' confirmation button and door open delay will be installed on the affected cart washers. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Device Technologies is providing work around instructions and will be upgrading affected Reliance 130L Cart Washers. |
Contact Information |
1800 429 551 - Device Technologies Regulatory Affairs |