Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01144-1 |
Product Name/Description |
DRX-1 System Console (radiation therapy digital imager)
All batch / serial numbers
ARTG Number: 161988 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
13/11/2012 |
Responsible Entity |
|
Reason/Issue |
Carestream Health Australia has received advice that there is a remote possibility that the DRX-1 System console has reversed cable neutral and ground wires on the interface box appliance connector. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Carestream is advising end users not to use potentially affected devices until they have been inspected by their technicians and undergone corrections, if necessary. |
Contact Information |