Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01142-1 |
Product Name/Description |
Access hLH Calibrators used with Access Immunoassay Systems. An in vitro diagnostic medical device (IVD)
Multiple Lot Numbers |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/11/2012 |
Responsible Entity |
|
Reason/Issue |
The affected Access hLH Calibrator lots do not meet the stated expiration date claim. As a result, the quality control and patient sample results generated while using these calibrator lot numbers might be falsely elevated by as much as 13%. This can occur up to four months prior to the expiration date listed on the calibrator vial. |
Recall Action |
Recall |
Recall Action Instructions |
Beckman Coulter is advising users to discontinue the use of the affected calibrator and is providing updated work around instructions to mitigate the issue. Clinicians/research scientists should be notified regarding potentially affected results at the discretion of the Laboratory Director. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |