Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01142-1
Product Name/Description Access hLH Calibrators used with Access Immunoassay Systems. An in vitro diagnostic medical device (IVD)

Multiple Lot Numbers
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/11/2012
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue The affected Access hLH Calibrator lots do not meet the stated expiration date claim. As a result, the quality control and patient sample results generated while using these calibrator lot numbers might be falsely elevated by as much as 13%. This can occur up to four months prior to the expiration date listed on the calibrator vial.
Recall Action Recall
Recall Action Instructions Beckman Coulter is advising users to discontinue the use of the affected calibrator and is providing updated work around instructions to mitigate the issue. Clinicians/research scientists should be notified regarding potentially affected results at the discretion of the Laboratory Director.
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre