Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01140-1
Product Name/Description AU5800 diluted detergent tank used with AU5800 Clinical Chemistry Analyser. An in vitro diagnostic medical device (IVD)

Multiple models and catalogue numbers

All serial numbers affected
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/11/2012
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue If the AU5800 diluted detergent tank valve is not positioned to 'CLOSED' during the instrument 'measure' mode, diluted detergent will not be dispensed and may adversely affect washing of cuvettes, probes and mixing-bars. Inadequate washing of cuvettes, probes and mixing bars can cause falsely high or low results to be reported for a variety of tests.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is providing work around instructions to mitigate the issue and is asking customers to add a label to the diluted detergent tank identifying the "Open" and "Closed" orientation of the valve.
Contact Information 1800 060 881 - Beckman Coulter Customer Support