Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01138-1
Product Name/Description ADVIA Centaur/Centaur XP/ Centaur CP HBcTotal (HBcT) Kits. An in vitro diagnostic medical device (IVD)

Siemens Material Number: 10309508

Product Number: 07566733

Multiple lot numbers affected

ARTG Number: 93126
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/11/2012
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics has received complaints regarding a perceived increased rate of reactive results that do not correspond with other HBV serological markers. Reagent lots ending in 044, 045 and 046 do not meet the specificity claims of 99.75% (5130/5143) with a 95% confidence interval (CI) of 99.57 to 99.87% as described in the Instructions for Use (IFU) for the HBcT assay when used on ADVIA Centaur systems.
Recall Action Recall
Recall Action Instructions Discontinue and discard affected lots. Affected lots will be replaced by Siemens Ltd.
Contact Information 1800 310 300 - Siemens Technical Support Centre