Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01135-1
Product Name/Description Multiple assays used with ADVIA Centaur, ADVIA Centaur XP, ADVIA Centaur CP Systems: Theophylline 2, Procalcitonin, Folate, DHEA-SO4, CA 19-9, BR Assay for CA 15-3, BR Assay for CA 27.29, BNP (B-type Natriuretic Peptide), In vitro diagnostic medical devices (IVDs)

Theophylline 2
ARTG: 175699

Procalcitonin, DHEA-SO4
ARTG: 175698

Folate
ARTG: 175700

BR Assay for CA 15-3, CA 19-9, BR Assay for CA 27.29
ARTG: 175636

BNP
ARTG: 175075

Multiple SN and REF Numbers
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/11/2012
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue The operating range of the ADVIA Centaur systems is 18°C to 30°C. The affected assays can product inaccurate results when performed within this operating range. A change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
This phenomenon is observed when assay results are reported at a temperature different from the temperature at which the assay was calibrated. Depending on the frequency of quality control testing and laboratory temperature at the time quality controls are tested, this issue may not always be detected through routine quality control monitoring.
Recall Action Recall for Product Correction
Recall Action Instructions Customers are being advised to immediately implement the revised room temperature operating ranges for the ADVIA Centaur, ADVIA Centaur XP, and ADVIA Centaur CP assays
Contact Information 1800 310 300 - Siemens Customer Care Centre