Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01112-1
Product Name/Description FlexiView 8800 Mobile C-Arms (Mobile, general-purpose, diagnostic, fluoroscopic x-ray system)

ARTG Number: 92396
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 31/10/2012
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare Surgery has finalised the corrections to remedy the issues communicated to the customers
in 2006 and 2007 (TGA Recall Reference #RN-2007-0077) . Specifically, these corrections will address issues involving system lock-ups,
incorrect image display, system stalls during the boot cycle, and failure of the system to display the air kerma and cumulative air kerma.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is providing work around instructions to mitigate the risk and will be implementing a software update to permanently correct the issue.
Contact Information 1800 659 465 - GE Customer Care Centre