Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01087-1
Product Name/Description Nihon Kohden Defibrillator (semi-automated defibrillator)

Model numbers: TEC 5521 & TEC 5531

ARTG Number: 172594
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 21/11/2012
Responsible Entity AHC Solutions Pty Ltd
Reason/Issue When the defibrillator operates on battery and the remaining battery power is almost zero, the battery mark becomes '0', the charge battery message appears and an alarm sounds. If this occurs and the battery isn't replaced with a charged battery or the defibrillator is not connected to AC power, the mother board might become damaged and the defibrillator may not work.
Recall Action Recall for Product Correction
Recall Action Instructions A software update will be implemented for the affected devices.
Contact Information 02 9980 9669 - AHC Solutions