Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01081-1
Product Name/Description Resuscitation Systems with blender field upgrade installed in GE Giraffe and Panda Warmers (infant radiant warmer)

ARTG Number: 139290
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 23/10/2012
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has recently become aware of a potential safety issue wherein the Oxygen and Air wall inlet fittings and/or labels on the back panel, were reversed during assembly. As a result, this may potentially have reversed Air/Oxygen mixer concentrations; for example, a setting of 100% Oxygen could have an output of 21% Oxygen, and vice versa. The settings of the blender knob will no longer be accurate.
Recall Action Recall for Product Correction
Recall Action Instructions If the affected unit is in use with a patient, GE is asking customers to disconnect the wall supply gas and switch to tank supply. If the affected unit is not currently in use, GE is asking customers to take the unit out of service. A GE Field Engineer will be deployed to the affected site to complete the required repair.
Contact Information 1300 722 229 - GE National Service and Support Centre