Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01073-1
Product Name/Description Elecsys Thyroglobulin assay used with the Elecsys Modular Analytics analysers An in vitro diagnostic medical device (IVD)

Material Number: 05118921-190

Lot Number: 16934601

Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/10/2012
Responsible Entity Roche Diagnostics Australia Pty Limited
Reason/Issue Roche have confirmed a positive bias of approximately 0.6 ng/mL with specific lots of the Elecsys Thyroglobulin assay
Recall Action Recall for Product Correction
Recall Action Instructions Affected laboratories to be aware of the potential elevated Tg recovery using this reagent lot and closely monitor internal quality control results.
Contact Information 1800 645 619 - Roche National Support Centre