Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01073-1 |
Product Name/Description |
Elecsys Thyroglobulin assay used with the Elecsys Modular Analytics analysers An in vitro diagnostic medical device (IVD)
Material Number: 05118921-190
Lot Number: 16934601
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/10/2012 |
Responsible Entity |
|
Reason/Issue |
Roche have confirmed a positive bias of approximately 0.6 ng/mL with specific lots of the Elecsys Thyroglobulin assay |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Affected laboratories to be aware of the potential elevated Tg recovery using this reagent lot and closely monitor internal quality control results. |
Contact Information |
1800 645 619 - Roche National Support Centre |