Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-01070-1 |
Product Name/Description |
Nihon Kohden Defibrillator CardioLife, Model Number: TEC 5521 & 5531 (Semi-automated defibrillator)
Serial Numbers: 01590, 00043 & 02082
ARTG Number: 116745 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
29/10/2012 |
Responsible Entity |
|
Reason/Issue |
The defibrillator will cease to work on DC power supply in a specific case (due to electrical damage to a protection circuit) when the battery is fully depleted and no longer able to hold a charge. There is no limitation for the operation of the defibrillator when the AC power is connected. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Quarantine and return affected devices to Boab Medical for a software update. |
Contact Information |
03 9809 4865 - Boab Medical Technologies |