Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01070-1
Product Name/Description Nihon Kohden Defibrillator CardioLife, Model Number: TEC 5521 & 5531 (Semi-automated defibrillator)

Serial Numbers: 01590, 00043 & 02082

ARTG Number: 116745
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 29/10/2012
Responsible Entity Dancor Pty Ltd T/a Boab Medical Technologies
Reason/Issue The defibrillator will cease to work on DC power supply in a specific case (due to electrical damage to a protection circuit) when the battery is fully depleted and no longer able to hold a charge. There is no limitation for the operation of the defibrillator when the AC power is connected.
Recall Action Recall for Product Correction
Recall Action Instructions Quarantine and return affected devices to Boab Medical for a software update.
Contact Information 03 9809 4865 - Boab Medical Technologies