Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01068-1
Product Name/Description Various mass spectrometers - Quattro micro, ACQUITY TQD, Quattro Premier XE, Xevo TQ; In vitro diagnostic medical devices (IVD's)

Quattro micro
Part:VB-QMicro
Serial Numbers:QAA1167, QAA412, QAA742, QAA65, QAA765, QAA802 & QAA836

ACQUITY TQD
Part:186003875
Serial Numbers:QBA470, QBB795, QBB924 & QBB928

Quattro Premier XE
Part:186003250
Serial Numbers:VAB1000, VAB1215, VAB610, VAB844, VAB845 & WAB998

Xevo TQ
Part:186004490
Serial Numbers: VBA626 & VBA744

ARTG Number: 201727
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/10/2012
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue There is a possibility that the ion block source heater may fail. In some instances, the failure could cause short term loss or variation of sensitivity that may subsequently lead to inaccurate data. The loss of or variation in sensitivity is most pronounced in assays with shorter dwell times and may not always be detected by quality control and internal standard value monitoring.
Recall Action Recall for Product Correction
Recall Action Instructions Customers to contact their Waters field service engineer to order a replacement part (ion block source heater).