Type of Product Medical Device
TGA Recall Reference RC-2012-RN-01047-1
Product Name/Description StreamLAB Automation Solutions (Clinical Chemistry Sample Processors). An in-vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/10/2012
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics has received reports from customers that the small pins at the bottom of the centre door panel of the Input/Output Module, which protrude approximately 5/8 inch, have contributed to a trip and fall incident. Customers are being asked to use caution when working in the area of the Input/Output Module and to post the communication near or on the StreamLAB Analytical Workcell system.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing work around instructions for users to implement.