Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00998-1 |
Product Name/Description |
IMMULITE /IMMULITE 1000 Osteocalcin Control Module (Clinical chemistry hormones). An in vitro diagnostic medical device (IVD)
Catalogue Number: LONCM
Siemens Material Number: 10385415
Control Lot Number: 0022
Kit Lots: 313 and above
ARTG Number: 179720 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
26/09/2012 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has identified an increase in the mean values obtained with the LONCM control lot 0022 when used with IMMULITE/IMMULITE 1000 Osteocalcin (LKON1, SMN 10361396) kit lots 313 and above. Siemens is providing revised control targets and ranges. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is providing work around instructions for users to implement. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |