Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-00996-1
Product Name/Description MOSAIQ (Radiation Oncology Information System). All releases pre version 2.41

ARTG Number: 180498
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/10/2012
Responsible Entity Elekta Pty Ltd
Reason/Issue A patient was mistreated after a user made and saved an inadvertent change to the definition of a treatment field in MOSAIQ. The mistreatment occurred at a site that had disabled the Field Edit Display feature.
Recall Action Recall for Product Correction
Recall Action Instructions Elekta is providing instructions to enable the Field Edit Delta Display Option and recommends upgrading to software version 2.41 or higher
Contact Information 02 8907 1800 - Elekta Regulatory Affairs