Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00992-1 |
Product Name/Description |
ARCHITECT HAVAb-IgG Assay (Indicated as an aid in the diagnosis of hepatitis A viral infection or detection of IgG anti-HAV ). An in-vitro diagnostic medical device
List Numbers: 6C29-20 and 6C29-25
Multiple Lot Numbers |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
26/09/2012 |
Responsible Entity |
|
Reason/Issue |
The ARCHITECT HAVAb-IgG assay shows a shift in the serum and plasma results to higher S/CO values over time, potentially leading to reduced specificity and increased false reactive results. The expiration dates for all affected reagent lots are being revised. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Abbott is providing work around instructions for users to implement until permanent preventive measures are implemented. |
Contact Information |
1800 816 696 - Abbott Customer Support |