Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00936-1 |
Product Name/Description |
Hospira LifeCare 4200 Patient-Controlled Analgesia (PCA) Infuser, PCA with MedNet Software Infusers (general purpose infusion pump).
4200 PCA Infuser; List Number 610F78201
PCA and PCA with MedNet Software Infusers; List Number: 612070901
ARTG Number: 138109 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
28/09/2012 |
Responsible Entity |
|
Reason/Issue |
Hospira has received reports overseas from customers related to the improper use of the slide clamp when the syringe (vial) was manipulated or when stopping infusion during delivery of fluids. Users are being informed of the correct usage of the slide clamp and/or syringe manipulation to prevent over delivery. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Customers are being advised of the correct use of the c slide clamp and/or syringe manipulation to prevent over delivery. |
Contact Information |
1300 046 774 - Option 1, Hospira Customer Service |